Modeling, Sensing and Control of Underactuated Continuum Robots for Endoluminal Navigation
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Cerebrovascular diseases, including stroke and aneurysm rupture, represent some of the most prevalent and morbid ailments experienced across the globe. Improvements in endovascular technology, including catheters and guidewires, have made minimally-invasive intervention the standard-of-care for these diseases. While neuroendovascular intervention provides lower risk than open surgery and higher efficacy than therapeutics alone, it is one of the most technically-demanding disciplines which is limited to highly-trained specialists at major hospitals. The burden placed on these experts is twofold: safely manipulating an inherently underactuated device throughout tortuous anatomy, as well as perceiving that device within the anatomy from biplane fluoroscopy alone.
This dissertation aims to address some key challenges of manipulation and perception as they relate to neuroendovascular intervention. The main technology underpinning the contributions of this dissertation is a robotically-actuated steerable microcatheter with distal articulation. This device enhances the dexterity beyond existing microcatheters while still enabling endoluminal passage of guidewires, aneurysm coils, or stent-retrievers. The first contribution of this dissertation involves the shape and pose estimation of steerable devices under biplane fluoroscopy. While extrinsic sensing modalities fail to scale down to the size constraints of thin-walled microcatheters, existing segmentation approaches fail to capture the full three-dimensional shape and pose, which is necessary for safe distal articulation. The next contribution leverages shape and pose information from biplane fluoroscopy to semi-autonomously steer the catheter in clinically-relevant phantom anatomy. In this approach, the interventionalist chooses a desired trajectory along the anatomy and controls insertion of the catheter, while the robot autonomously steers the underactuated device to best follow the trajectory. The final contribution introduces a sensory technology specifically designed for treating acute ischemic stroke by mechanical thrombectomy. This system detects when the catheter comes into contact with a blood clot without relying on any distal sensing instrumentation, and has been shown to improve experts’ perception compared to fluoroscopy alone.
The contributions of this dissertation combine elements of modeling, sensing, and control of flexible robotic catheters to address manipulation and perception challenges experienced during neuroendovascular intervention. We believe they demonstrate improvements which may elevate the standard-of-care, and in doing so may improve the accessibility of cerebrovascular treatment.